Equipment/Cardiology/F1 and F2
Automated external defibrillator
Comen F1 and F2
Automated external defibrillators for adult and pediatric patients, in semi-automatic and fully automatic versions, with voice guidance and a disposable battery.
The buttons on this page open a WhatsApp chat at +55 61 99321-6835 with the model name already filled in. Our AI answers questions about the equipment; for the full datasheet, quotes, technical visits or support, the São Bernardo team takes over the chat. Hours: Monday to Friday, 8 am to 5:30 pm.
Prefer not to use WhatsApp? Call +55 61 3233-0536 or +55 61 3233-2958, or email contato@saobernardodf.com.br.

Spec sheet
Key specifications
- Patients
- Adult and pediatric, with a patient selector switch and electrode pads that identify the type automatically
- Operation
- Semi-automatic and fully automatic versions
- Energy
- BTE waveform with patient impedance compensation; adults 100 to 360 J and children 10 to 100 J
- User guidance
- Voice prompts and voice CPR guidance
- Battery
- Disposable, non-rechargeable, maintenance-free; more than 6 years on standby
- Storage
- Up to 5 hours of ECG, 1,000 alarm events, and 1,500 self-test reports
- Communication
- Wi-Fi and cellular network (2G/3G/4G/5G)
- Weight and dimensions
- 2.0 ± 0.2 kg with battery and without network module; 295 × 210 × 75.6 mm
- Protection
- IP55
Models
What differs between them
| Item | F1 | F2 |
|---|---|---|
| Display | No display | 7-inch TFT, 800 × 480, with one ECG trace |
| ECG | Not applicable | Via the multifunction electrode pads, leads I, II, and III |
| Monitoring run time | At least 15 hours | At least 12 hours |
| 150 J shocks per battery | At least 520 | At least 460 |
| 200 J shocks per battery | At least 410 | At least 360 |
F1
- Display
- No display
- ECG
- Not applicable
- Monitoring run time
- At least 15 hours
- 150 J shocks per battery
- At least 520
- 200 J shocks per battery
- At least 410
F2
- Display
- 7-inch TFT, 800 × 480, with one ECG trace
- ECG
- Via the multifunction electrode pads, leads I, II, and III
- Monitoring run time
- At least 12 hours
- 150 J shocks per battery
- At least 460
- 200 J shocks per battery
- At least 360
Manufacturer data: Comen, F1/F1 FullyAutomatic AED Specification, V1.2, 2025-06-18; Comen, F2/F2 FullyAutomatic AED Specification, V1.2, 2025-06-17. Configurations and optional modules vary, and specifications may be changed by the manufacturer. Confirm the configuration with São Bernardo.